Work Instruction for Sartorius & Mettler balance Operation, Calibration and Maintenance of Analytical balance

Work Instruction


Operation, Calibration and Maintenance of Analytical balance (Sartorius)

                                                         (Analytical Science) 


Lab Code



Version No.:                        
Effective Date:                    

Review Date:                                  

Supersedes:                       

Author(s):                                        




Technical Approver
Authorized By
Name



Designation



Signature




Date




1.     Scope
       This work instruction is applicable for the operation; calibration and maintenance of the following    
       Analytical balance located in Laboratory.

S. No.
Make
Model
Readability
1
Sartorius
GP3202-OCE
0.5g – 3200g

2.     Procedure
2.1       Operation:
2.1.1     Ensure leveling of the balance by turning the two leveling feet at the rear of the                        balance   housing until the air bubble is in the centre of the leveling control.
2.1.2       Press the “І/Φ” key to switch on the balance. When the normal weight display appears the balance is ready for weighing. Press the container / butter paper on the weighing pan.
2.1.3     Press “TARE” key to tare the weight.
2.1.4     When the zero display appears the balance is ready for weighing.
2.1.5     Transfer the weighing sample on the pan/container.
2.1.6     When the balance/scale is switched on, the “         “symbol is displayed until you press the   
            key, If the symbol “          ” displayed during the operation, this indicates that the processor
            is performing a function and cannot receive further command at the moment.
2.1.7       Now read off the displayed weight.
2.1.8     Switching between the weighing units
 The balance displays the result in weighing unit 1 (say g)
•Press the “F” key for two seconds.
•The balance displays the result in weighing unit 2 (say kg)
•To return the weighing unit 1, press the “F” key again.
2.1.9     To switch off the balance, press and hold the “І/Φ” key until the message “OFF” appears in
             the display.
2.1.10    After switching off, the balance will be in standby mode. For another weighing to be perfomed, place the weighing sample on the pan and the balance will display the result immediately.
            Note:  it is advisable not to disconnect the instrument from the power supply but to switch it    
            off only by using the “І/Φ” key which assures that the balance is always in thermal equilibrium.



2.1.11    Refer to the manufacture’s manual for further guidance on the operation of the controls of individual instruments.
2.1.12    Avoid the use of weighing vessels made from ferrous metals on analytical balance, in order to reduce the effects of magnetic fields on the readings. For the same reason, magnetic stirrer bars must not be placed in vessels whilst weighing.
2.1.13    Avoid the use of plastic vessels on analytical balance in order to reduce the effects of static electricity on the readings.
2.2         Calibration and Frequency:
2.2.1     Balance is calibrated externally by a service engineer once in every 4 months for the
            following parameters.
·               Linearity
·               Corner Load Error
·               Measurement Uncertainty
2.2.2     For external calibration ensure a calibration certificate is provided along with the traceability
             certificates of the standards used in calibration.
            Note: All weights used in calibration must be traceable to national or international standards.
2.2.3     Linearity: Record the readings of two calibration weights. These should represent
            approximately 25% and 75% of the instrument’s capacity, where deviation from ideal linear
            behaviour is greatest.  In addition, full range calibration may be performed covering minimum
            to maximum of the   balance capacity.
2.2.4     Acceptance criteria: The readings must be within the tolerances determined for the certified   
            weights, e.g. those provided in the table below based on ±0.1% as per USP norms.
2.2.5       Corner Load Error: Place a weight equivalent to approximately 25% of the instrument’s capacity on the pan at the following locations: - Centre, Left Front, Left Rear, Right Rear, Right Front. Record the reading at each location.
2.2.6       Acceptance Criteria:  Each reading must meet the acceptance criteria stated under linearity.
2.2.7       Measurement Uncertainty: Place selected mass on the pan for 10 times and record the reading each time. Calculate the standard deviation (SD) for the 10 readings.
2.2.8       Calculate the minimum weight as (3  x  SD) / Nominal Weight
2.2.9       Acceptance criteria: <0.001g of the amount weighed

2.3        Calibration :
2.3.1     If any of the calibration tests are outside their specified tolerance, follow the
            procedure outlined below.
2.3.2       Determine if there is a reason for the failure. If a reason is found which can be corrected, make any necessary adjustments and repeat the tests.
2.3.3       If the tests pass, the original data is invalidated and the instrument can continue to be used.
2.3.4       If no reason for the failure can be determined, perform the tests twice. If both pass the instrument can continue to be used.
2.3.5       If any of the re-tests fail, identify the balance as defective and label it “Out of Service” and contact the service engineer to arrange a service visit.
2.4          Maintenance:
2.4.1     Cleaning
2.4.1.1  Remove the weighing pan and weighing chamber plate.
2.4.1.2  Clean the weighing chamber with a soft brush or a damp cloth.
2.4.1.3  Commercially available cleaning agents may be used for cleaning the balance. Do not use                
            abrasive cleaners or powerful solvents.
2.5        Servicing:
2.5.1.     Regular servicing of the equipment is covered under annual maintenance contract which ensures smooth functioning of the balances.

2.6        Special Instructions
     Nil

2.7        Reference

           Operating instructions, Sartorius AG


3.0  Appendices
         Nil

4.0  Revision History

WI No.
Supersedes No. & Date
Changes made (in brief)





============================================================================
Work Instruction

Operation, Calibration and Maintenance of
Analytical Balance 
(Mettler)

                                                         (Analytical Science)

Lab code…


Version No.:                                    

Effective Date:                    

Review Date:                                  

Supersedes:                       
Author(s):                                        



Technical Approver
Authorized By
Name



Designation



Signature




Date





1.            Scope
This work instruction is applicable for the operation, calibration and maintenance of the following analytical balances located in the Chemical & microbiology labs of Analytical Sciences.

S. No
Make
Model
Range
1
Mettler Toledo
AG 245
0.01 mg to 41 g
0.1 mg to 210 g
2
Mettler Toledo
AG 204
0.1 mg to 210 g
     3
Mettler Toledo
AB 304-S/FACT
10 mg to 320 g

2.            Procedure
2.1.         Operation:
2.1.1.     Ensure levelling of the balance by turning the two levelling feet at the rear of the balance housing until the air bubble is in the centre of the levelling control.
2.1.2.     Press the << ON/OFF >> key to switch on the balance. When the normal weight display appears, the balance is ready for weighing.
2.1.3.     Open the right draft shield door in Mettler Toledo balance.
2.1.4.     Place the container/ butter paper on the weighing pan and close the shield door. (Ensure all the doors are closed prior to weighing each time to get accurate weighing results)
2.1.5.     Press << → O/T ← >>key to tare the weight.
2.1.6.     When the zero display appears the balance is ready for weighing
2.1.7.     Open the draft shield, transfer the weighing sample on the pan and close the draft shield
2.1.8.     Wait until the circular symbol of the stability detector fades. When the symbol has faded, the weighing result is stable
2.1.9.     Now read off the displayed weight
2.1.10.   Switching between weighing units
·         The balance displays the result in weighing unit 1 (say g)
·         Press the << F >> key briefly
·         The balance displays the result in weighing unit 2 (say mg)
·         To return to weighing unit 1, press the << F >> key again
2.1.11.   To switch off the balance, press and hold the  << ON/OFF >> key until the message “OFF” appears in the display
2.1.12.   After switching off, the balance will be in the standby mode. For another weighing to be performed, place the weighing sample on the pan and the balance will display the result immediately. (No need to switch it on using the << ON/OFF >> key)
Note: It is advisable not to disconnect the instrument from the power supply but to switch it off only by using the << ON/OFF >> key which assures that the balance is always in thermal equilibrium. 
2.1.13.   Model AG-245 is a dual-range balance. This has a fine (semi micro) range from 0 to 41 grams. In this fine range the balance shows the result with a higher resolution, i.e. with one decimal place more. This fine range cannot be moved, i.e. it always starts at 0 and ends always at 41 grams.  The balance automatically operates in the normal weighing range when first switched on. By briefly pressing the «1/10d» key, you can switch to the fine range. The fine range remains active up to a weight of 41 grams. 

Note: Below 41 grams, you can switch between the fine range and the normal weighing range at any time by pressing the «1/10d» key. If the weight is greater than 41 grams, the balance quits the fine range and displays in the normal weighing range. If you remove or decrease the weight following a weighing in the range above 41 grams, the balance automatically returns to the fine range.

2.1.14.   Refer to the manufacturer’s operating instructions for further guidance on the operation of the controls of individual instruments.
2.1.15.   Avoid the use of weighing vessels made from ferrous metals on analytical balances, in order to reduce the effects of magnetic fields on the readings. For the same reason, magnetic stirrer bars must not be placed in vessels whilst weighing.
  Note: Avoid the use of plastic vessels on analytical balances in order to reduce the effects of   
  static electricity on the readings.

2.2.         Calibration and Frequency:
2.2.1.     Balance is calibrated internally (2 weights check every day) and externally by a service engineer once in a year for the following parameters:
·              Linearity
·              Corner Load Error
·              Measurement Uncertainty
Note: Daily calibration is not necessary if balance is not used for the day
2.2.2       Record the internal calibration readings in the log as per Appendix 1 and for external calibration ensure a calibration certificate is provided along with the traceability documents.
Note: All weights used in calibration must be traceable to national or international standards.
2.2.3       Linearity: Record the readings of two calibration weights. These should represent   
approximately 25% and 75% of the instrument’s capacity, where deviation from ideal linear behaviour is greatest. In addition, full range calibration may be performed covering minimum to maximum of the balance capacity.
S. No
Make
Model
Range
Daily Calibration check Weights
1
Mettler Toledo
AG 245
0.01 mg to 41 g
0.1 mg to 210 g
         10g & 31g   
          50g & 160g 
2
Mettler Toledo
AG 204
0.1 mg to 210 g
          50g & 160g 
     3
Mettler Toledo
AB 304-S/FACT
10 mg to 320 g
          80g & 240g

2.2.4       Acceptance criteria: The readings must be within the tolerances determined for the certified weights, e.g. those provided in the table below based on ±0.1% as per USP norms.

Nominal Weight
Allowable Range (g)
Nominal Weight
Allowable Range (g)
1mg
0.000999 - 0.001001
50g
49.95  - 50.05
10mg
0.00999 - 0.01001
100g
99.90 - 100.10
100mg
0.0999 - 0.1001
160g
159.84 - 160.16
500mg
0.4995 - 0.5005
200g
199.80 - 200.20
1g
0.999 - 1.001
400g
399.6 -  400.4
10g
9.99 - 10.01
500g
499.5 - 500.5
20g
19.98 - 20.02


2.2.5       Corner Load Error: Place a weight equivalent to approximately 25% of the instrument’s capacity on the pan at the following locations: - Centre, Left Front, Left Rear, Right Rear, Right Front. Record the reading at each location.
2.2.6       Acceptance Criteria:  Each reading must meet the acceptance criteria stated under linearity.
2.2.7       Measurement Uncertainty: Place selected mass on the pan for 10 times and record the reading each time. Calculate the standard deviation (SD) for the 10 readings.
2.2.8       Calculate the minimum weight as (3  x  SD) / Nominal Weight
2.2.9       Acceptance criteria: <0.001g of the amount weighed

2.3          Calibration or Performance Check Failure
2.3.1       If any of the calibration tests are outside their specified tolerance, follow the procedure outlined below.
2.3.2       Determine if there is a reason for the failure. If a reason is found which can be corrected, make any necessary adjustments and repeat the tests.
2.3.3       If the tests pass, the original data is invalidated and the instrument can continue to be used.
2.3.4       If no reason for the failure can be determined, perform the tests twice. If both pass the instrument can continue to be used.
2.3.5       If any of the re-tests fail, identify the balance as defective and label it “Out of Service” and contact the service engineer to arrange a service visit.
2.4          Maintenance:
2.4.1       Cleaning
2.4.1.1   Remove the weighing pan and weighing chamber plate.
2.4.1.2   Clean the weighing chamber with a soft brush or a damp cloth.
2.4.1.3   Commercially available cleaning agents may be used for cleaning the balance. Do not use abrasive cleaners or powerful solvents.

2.4.2       Servicing:
2.4.2.1   Regular servicing of the equipment is covered under annual maintenance contract which ensures smooth functioning of the balances.
         
3.         Appendices
  Appendix 1: Internal Calibration Log for Linearity Test in Analytical Balance (xxx)